  Datasheet File OCR Text:
  the mbc 275 series of ultra low open frame medical power supplies feature a wide universal ac input range of 85 v C 264 vac, offering 275 w of output power with forced air cooling in a variety of isolated single output voltages. the mbc series was designed to 3r d edition medical approvals and provides 2x mopp (means of patient pr otection) isolation for class i and class ii installations. these power supplies are ideal for medical, t elecom, datacom, industrial equipment and other applications. ? 5 x 3 x 0.75 inches form factor ? 275 watts with forced air cooling ? approval to en60601 3rd edition ? efficiencies up to 92% ? - 40 to 70c degree o perating temperature ? dual fusing ? 12 v / 0.5 a fan output, thermal shut - down feature ? 3.37 m illion hours, telcordia - sr332 - issue 3 mtbf ? standby power < 0.5 w ? medical (bf) safety approvals ? diagnostic ? drug pump ? monitoring ? dialysis ? home health care ? portable equipment
2 m bc 275 series tech.support@psbel.com mbc275 - 1t12l screw terminal 12 v 12.5 a 13.33 a 20.83 a 0.0 a 2 % mbc275 - 1012l molex connector mbc275 - 1t15 l screw terminal 15 v 10 a 10.66 a 16.67 a 0.0 a 2 % mbc275 - 1015l molex connector mbc275 - 1t 24 l screw terminal 24 v 6.25 a 6.67 a 10.41 a 0.0 a 1% mbc275 - 10 24 l molex connector mbc275 - 1t 30 l screw terminal 30 v 5 a 5.33 a 8.33 a 0.0 a 1% mbc275 - 10 30 l molex connector mbc275 - 1t48l screw terminal 48 v 3.12 a 3.33 a 5.2 a 0.0 a 1% mbc275 - 1048l molex connector mbc275 - 1t 58 l screw terminal 58 v 2.58 a 2.76 a 4.31 a 0.0 a 1% mbc275 - 1058l molex connector cover - 275 - xbc metal cover kit accessory 1 ripple is peak to peak with 20 mhz bandwidth and 10 f (tantalum capacitor) in parallel with a 0.1 f capacitor at rated line voltage and load ranges. specifications are for nominal input voltage, 25c unless otherwise stated. input voltage universal (derate from 100% at 100 v ac to 72% for forced c ooling and 69% for convection c ooling at 80 v ac) 8 0 - 264 vac / 390 vdc input frequency 47 - 63 hz input current 115 vac: 230 vac : 2.6 a max. 1.3 a max. no load power typical for mbc275 - 1xxx typical for mbc275 - 1xxx - pgpf < 0.5 w < 0.85 w inrush current 115 vac: 230 vac: 264 vac: 25 a 45 a 75 a leakage current typical (n.a. for class ii option - without input earth pin ) touch current 3 00 ua < 1 00 ua power factor at full load > 0.95 switching frequency pfc pwm 70 to 130 k hz 50 to 80 k hz
m bc 275 series 3 asia - pacific +86 755 298 85888 europe, middle east +353 61 225 977 north america +1 408 785 5200 ? 2016 bel power solutions & protection bcd.00 88 2 _001 output power 2 with 13 cfm forced air cooling with natural convection cooling at 100 to 264 vac 275 w up to 16 0 w efficiency ( typical @ 230 vac full load ) 48 v , 58 v : 24 v , 30 v : 12 v, 15 v: 92 % 9 0 % 88 % hold - up time at 275 w: at 16 0 w: 8 ms 16 ms line regulation +/ - 0.5% load regulation +/ - 1% transient response 25% step load change, at 0.1 a/us slew rate, 50% duty cycle, 50 hz = 4% recovery time < 5 ms voltage adjustment +/ - 3% rise time typical 55 ms set point tolerance 3 +/ - 1% over current protection > 110% over voltage protection 110 to 140% short circuit protection hiccup mode 2 combined output power of main output, fan supply shall not exceed max. power rating. 3 fan supply output voltage tolerance including set point accuracy, line and load regulation is +/ - 10% and ripple and noise is less than 10%. operating temperature startup guaranteed with spec . deviation - 40 to +70c - 40 to 0c storage temperature - 40 to +85c relative humidity non - condensing 5% to 95% altitude operating: non - operating: 16,000 ft 40,000 ft. mtbf telcordia - sr332 - issue 3 3.37 million h ours conducted emissions en55022 - b, cispr22 - b, fcc part15 C
4 m bc 275 series tech.support@psbel.com isolation voltage input to output: ( medical applications ) input to gnd: (not applicable for class ii option) output to gnd: for type bf ( for type b (n/a f or class ii option) 4 000 vac 1500 vac 1500 vac 500 vac safety standard(s) approved to the latest edition of the following standards: csa/ul60601 - 1, en60601 - 1 and iec60601 - 1. agency approvals nemko, ul, c - ul ce mark complies with lvd directive ac input connector j1 pin 1 pin 2 pin 3 ac line not fitted ac neutral molex: 26 - 60 - 4030 mating: 09 - 50 - 3031; pins: 08 - 50 - 0106 dc output connector j2 pin 1 , 2, 3 pin 4. 5. 6 v 1 + ve v1 - ve option 1 (screw terminal): molex: 39357 series or equivalent option 2 (molex connector): m olex: 26 - 60 - 4060 mating: 09 - 50 - 3061; pins: 08 - 50 - 0106 aux (fan) output j3 pin 1 pin 2 fan + ve fan - ve amp: 640456 - 2 mating: 640440 - 2 signal output 4 j4 pin 1 pin 2 pin 3 vs pgpf gdn amp :640456 - 3 mating: 640440 - 3 4 for pgpf signal output connector option please contact factory. weight approx. 2 00 g dimensions 127 x 76.2 x 19.05 mm ( 5 x 3 x 0.75 inches ) cooling 275 w with 13 cfm forced air cooling (refer to mechanical drawing) up to 16 0 w with natural convection cooling (refer to derating curve)
m bc 275 series 5 asia - pacific +86 755 298 85888 europe, middle east +353 61 225 977 north america +1 408 785 5200 ? 2016 bel power solutions & protection bcd.00 88 2 _001 mechanical drawing C option 1 without pgpf mechanical drawing C option 2 without pgpf
6 m bc 275 series tech.support@psbel.com mechanical drawing C option 1 with pgpf mechanical drawing C option 2 with pgpf
m bc 275 series 7 asia - pacific +86 755 298 85888 europe, middle east +353 61 225 977 north america +1 408 785 5200 ? 2016 bel power solutions & protection bcd.00 88 2 _001 nuclear and medical applications - products are not designed or intended for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems. technical revisions - the appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. specifications are subject to change without notice.


▲Up To Search▲   

 
Price & Availability of COVER-275-XBC

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X